Sign In | Join Free | My burrillandco.com |
|
Ask Lasest Price | |
Brand Name : | Lonroy |
Model Number : | BGD 8140S |
Certification : | Includes Calibration Certification |
Price : | Negotaible |
Payment Terms : | L/C,D/A,D/P,T/T,Western Union |
Supply Ability : | 200 |
Delivery Time : | 5-8 work days |
BGD 8140S Multifunctional Irradiance Radiometer
1.Description
For all artificial accelerated light-aging equipment, irradiance (or radiant energy) is one of the most critical parameters
affecting test results. Over time, components such as filters,
irradiance probes, and sensors in aging test chambers may degrade,
leading to drift in displayed irradiance values. Therefore, regular
calibration of irradiance is essential to ensure accuracy in accelerated aging tests.
The BGD 8140S Multifunctional Irradiance Radiometer, developed by Biaogeda Company, is a revolutionary handheld
instrument designed for this purpose. It consists of a portable
irradiance calibration meter (main unit) and one or more
specialized probes. To use:
1. Select the appropriate probe based on the calibration target.
2. Insert the probe into the radiometer and choose the lamp type
(e.g., xenon or UV) and filter on the display.
3. Place the probe inside the aging chamber and activate the test
equipment.
4. Calibrate the irradiance by comparing the values displayed on the
radiometer with the chamber's readings.
Additionally, when a probe is connected, the radiometer
automatically reads and displays the probe's factory ID and
calibration date, ensuring users track recalibration schedules.
The BGD 8140S is compatible with all Biaogeda aging chambers (xenon arc and
fluorescent UV) and offers unique advantages over conventional radiometers:
* Automatic Probe Recognition: Identifies probe type and parameters
upon insertion.
* Universal Xenon Probe: Calibrates irradiance for both air-cooled
and water-cooled xenon lamps across multiple filters.
* UV Probe Versatility: Supports calibration of all fluorescent UV
lamps (UVA/UVB) with a single probe.
* High-Resolution Display: 5-inch capacitive touchscreen (1080×720)
for intuitive operation.
* Modular Design: Separate probes and main unit reduce costs compared
to integrated systems.
* Flexible Calibration Management: Probes can be recalibrated or
replaced after expiration.
* Bilingual Interface: Built-in Chinese and English language options.
2.Technical Parameter
Range | - 0.00~2.00 W/m² (340nm) - 0.00~4.00 W/m² (420nm) - 0~300 W/m² (300nm~400nm/TUV) - 0.00~2.00 W/m² (UVA&UVB) |
Irradiance Error | ±10% |
Recommended Calibration Cycle | One year |
Probe Temperature Drift | ±0.02%/℃ |
Operating Environment | 0~50℃; 10%RH~90%RH |
Maximum Working Temperature | 70℃ (Probe) |
Main Body Size | 143mm×75mm×20mm |
Net Weight | 380g (Main body and probe) |
![]() |